Liquid Controls M-15 - User Manual

Default Icon
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...

Other manuals for Liquid Controls M-15

Questions and Answers

  • D
    Danielle MeyerAug 16, 2025
    What to do if there is no flow through Liquid Controls Measuring Instruments?
    • E
      Eric BuckAug 17, 2025
      If your Liquid Controls Measuring Instruments meter is not flowing any product, there might be a few reasons. It could be due to a faulty or non-functioning pump, a valve that is not open or not working correctly, or the meter might be 'frozen' because of chemical 'salts' or foreign material buildup inside the metering chamber. Consider cleaning the meter and inspecting for any damage.
  • V
    Valerie MartinAug 19, 2025
    How to fix if Liquid Controls M-15 Measuring Instruments meter counts down in reverse?
    • K
      Kelsey HoltAug 20, 2025
      If your Liquid Controls Measuring Instruments meter counts down in reverse, you should reverse the direction of flow by reversing the adjuster drive gear.
  • J
    Juan HarveyAug 22, 2025
    Why is my Liquid Controls M-15 meter running too slowly?
    • P
      pgibsonAug 22, 2025
      If your Liquid Controls Measuring Instruments meter runs too slowly, several factors could be the cause. The valve's internal mechanism might be faulty, preventing it from fully opening. Also, the meter gears or rotors could be partially 'salted', which slows down the rotation of parts. Cleaning the meter might help. Another potential cause could be a downstream restriction or a clogged strainer basket.
  • R
    Richard JosephAug 25, 2025
    How to stop leakage from the cover gasket on Liquid Controls M-15?
    • L
      leslie44Aug 26, 2025
      If you observe leakage from the cover gasket of your Liquid Controls Measuring Instruments, it might be because the gasket has been damaged due to shock pressure, or the cover bolts haven't been tightened sufficiently. Replace the gasket and ensure the cover bolts are correctly tightened.
  • C
    cfieldsAug 28, 2025
    How to fix leakage past packing gland drive shaft housing in Liquid Controls Measuring Instruments?
    • R
      Raymond HillAug 28, 2025
      If you notice leakage past the packing gland drive shaft housing from the internal metering chamber in your Liquid Controls Measuring Instruments, it's likely that the internal seal of the packing gland assembly is worn. Replace the packing gland and O-ring seal.
  • H
    Haley CarterAug 30, 2025
    Why are teeth breaking on timing gears in Liquid Controls M-15 Measuring Instruments?
    • J
      Jesus FreemanAug 30, 2025
      If the teeth on the timing gears of your Liquid Controls Measuring Instruments are breaking, it could be due to starting or stopping the flow in the meter too rapidly, or the pump bypass not being adjusted properly.
  • K
    Kendra BakerSep 2, 2025
    How to fix the Liquid Controls M-15 Measuring Instruments when it counts down in reverse?
    • G
      Garrett GomezSep 2, 2025
      If the Liquid Controls Measuring Instruments meter counts down in reverse, reverse the direction of flow by reversing the adjuster drive gear.
  • J
    Jessica SmithSep 5, 2025
    What to do if there is leakage from the cover gasket of Liquid Controls M-15 Measuring Instruments?
    • K
      Katherine CastanedaSep 5, 2025
      If you observe leakage from the cover gasket of your Liquid Controls Measuring Instruments, it's likely that the gasket has been damaged due to shock pressure or the cover bolts haven't been tightened sufficiently. Replace the gasket and ensure the cover bolts are correctly tightened.
  • J
    Jonathan RamirezSep 7, 2025
    Why is there no flow through my Liquid Controls M-15 Measuring Instruments?
    • S
      Sara AvilaSep 7, 2025
      If the Liquid Controls Measuring Instruments has no flow, this could be due to a faulty non-functioning pump or a valve that is not open or not functioning.
  • D
    Daniel HarrisSep 12, 2025
    What to do if product flows through Liquid Controls Measuring Instruments but the register doesn't operate?
    • M
      Mary HillSep 12, 2025
      If product flows through the Liquid Controls Measuring Instruments meter but the register does not operate, check the packing gland and gear train. A faulty register should be checked and repaired by a trained mechanic. Also, replace the packing gland, as this may be caused by starting the flow too rapidly.

Summary

Safety Procedures

BE PREPARED

Safety preparation steps before using the meter, including proper training and precautions.

SAFELY EVACUATE PIPING SYSTEM

Procedures for safely evacuating the piping system before disassembly, ensuring zero pressure.

OBSERVE NATIONAL & LOCAL CODES

Adherence to electrical and safety codes for installations, ensuring compliance with regulations.

IN THE EVENT OF A GAS FIRE

Actions for handling gas fires, including evacuation and contacting the fire department.

Installation & Operation

How LC Meters Work

Explanation of the positive displacement meter mechanism and its operation.

Meter Start Up and Operation

Placing the Meter into Operation

Step-by-step procedure for filling the meter and system with liquid for start-up.

Maintenance

Liquid Controls M-15 Specifications

General IconGeneral
BrandLiquid Controls
ModelM-15
CategoryMeasuring Instruments
LanguageEnglish