M & MA Series Meters
46
Get the latest PDF manual:
https://www.lcmeter.com/resources/technical/manuals
Mobile/online version of this manual:
https://www.lcmeter.com/manuals
Product flows through meter
but register does not record
correctly
This may result from the following:
·
Adjuster not properly calibrated.
·
Incorrect gear plate or gear ratio has been
installed.
·
Air is in the system.
No flow through meter
This may result from the following:
·
Faulty non-functioning pump.
·
Valve not open or not functioning.
·
Meter frozendue to build-up of chemical salts”
or foreign material inside metering chamber.
Clean the meter and inspect for damage.
Meter runs too slowly
This may result from the following:
·
Valve internal mechanism faulty. Valve does not
open fully.
·
Meter gears or rotors partially “salted” enough to
slow up rotation of parts. Clean the meter.
·
Downstream restriction.
·
Clogged strainer basket.
The meter counts down in
reverse (the meter numbers
decrease)
Reverse the direction of flow by reversing the
adjuster drive gear. See Reversing the Meter
Registration .
18

Other manuals for Liquid Controls MA-7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Liquid Controls MA-7 and is the answer not in the manual?

Liquid Controls MA-7 Specifications

General IconGeneral
BrandLiquid Controls
ModelMA-7
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals