EasyManua.ls Logo

LKC TECHNOLOGIES RETeval - Page 2

LKC TECHNOLOGIES RETeval
2 pages
Print Icon
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LKC Technologies, Inc.
2 Professional Drive, Suite 222, Gaithersburg, MD 20879 USA
t: +1 301.840.1992 | e: sales@lkc.com | www.lkc.com
LKC Europe | Finland
t: +358 40 8486625 | e: europe@lkc.com | www.lkc.com
Reordering electrodes? Please visit LKC’s online store at store.lkc.com.
5
6
7
2. The screen will present opons prior to connecng the
Sensor Strip electrode to the Sensor Strip lead. “Cancel”
will cancel the enre test. “Done OD/OS” will skip all tests
of the specied eye. “Skip” allows the user to skip current
test step.
3. Connect the Sensor Strip electrode to the Sensor Strip
lead. The blue tab should face outwards. The device will
verify that your electrode is connected securely, and that
the electrode noise is suciently small enough to proceed.
Select “Next”.
BEGIN TEST
1. Place the eye cup of the device so that it rests evenly over
the paent’s right eye rst, if doing bilateral test. Inform
the paent to focus on the red light in the center of the
ganzfeld. Ensure the paents pupil is located within the
green circle shown on the screen. Select “Start test”.
Device will progress through each step.
TIP: For Td protocols, the whole pupil must be visible to
the operator. The test will not proceed if it is not visible.
Make sure to wait for the blue circle with red crosshatch
to appear over the pupil.
2. Upon compleon of a step, certain protocols will allow you
to review the waveform. If you are sased, select “Next”,
or select “Repeat” to repeat the step again.
3. Aer all steps are completed, the device will automacally
save the results.
4. You will have the opon to “Retest” the paent once
you’ve viewed their results
5. Once tesng is completed, disconnect the Sensor Strip
lead from the electrode. Remove electrodes from paent’s
face.
FILE REPORT
1. Connect the RETeval device’s docking staon to your
computer using the USB cable. Set the RETeval device in
the docking staon. On your computer, the RETeval device
will show similar to an external thumbdrive. To access the
PDF reports, locate the “Reports” folder, which is sorted
by date modied. The PDF report follows the format of
“PaentID_DOB_TimeStamp.
CLEAN DEVICE
1. Cleaning of the eyecup and Sensor Strip lead is
recommended between paent uses. The RETeval device
exterior can be cleaned with either 70% isopropyl alcohol
or alkyl dimethyl benzyl ammonium chloride.
Refer to user manual before use. RX only.
© 2020 LKC Technologies, Inc. US 92-078.008.001 5/2020
Part #: 96-047 5/20/2020

Other manuals for LKC TECHNOLOGIES RETeval