CONFIDENTIAL |LojerOy|Putajantie42,FI38210Sastamala,Finland|www.lojer.com
Ver. 1.0, 25/10/2022
Frame and moulded soft surfaces
• Cleantheframe and other hard surfaceswithadamp(micro)berclothandamilddetergent
solution(neutralpH6–8orweaklyalkalinepH8–10).Payspecialattentiontothorough
cleaningofcontactsurfaces.
• Cleanallmoulded soft surfaceswithadamp(micro)berclothandaneutraldetergent
solution(pH6–8).
>
Forexample,Desiplint(availableatwww.lojer.com)canbeusedtocleanand
disinfectthedevice.
>
Donotusee.g.anysolvents,abrasivecleaningsagentsorscouringpads
astheycandamagethesurfaces.
• Useasoftbrushtocleandifcultstains,cornersandotherhardtoreachplacesonhard
surfaces.
• Removedetergentresiduesorexcessdetergentbywipingthesurfaceswithacloth
dampenedwithcleanwater(followtherespectivedetergentmanufacturer’sinstructions).
• Allowthesurfacestofullydrybeforeusingorstoringthedevice.
Disinfecting
Thesurfacesshouldalwaysbecleanedbeforedisinfection.Usedisinfectantonlyifjustied
(e.g.topreventthetransmissionofharmfulmicrobes)asdisinfectionagentsmightchange
thesurfacestructureofmaterialsovertime.
• Allsecretionstainsshouldberemovedimmediatelyaftertheyappear.
• Followthedisinfectinginstructionsgivenbythedisinfectantmanufacturers.
• Anysurfaceshouldnotbesubjecttolong-termexposurebyanytypeofliquids.
All surfaces
• Disinfectthesurfaceswithadamp(micro)bercloth,usingdisinfectantssuitablefordisinfecting
medicaldevices,inaccordancewiththerespectivemanufacturer’sintendedpurposeand
instructionsforuse.
>
Forexample,max.70%solutionofethanolorpropanolcanbeusedtodisinfect
secretionstains.
• Wipethesurfaceswithaclothdampenedwithcleanwater.
• Allowthesurfacestofullydrybeforeusingorstoringthedevice.
Suitable tested disinfectants:
Alcohols Ethanolmax.70%
Propanol60%
BacillolAF
Terralin
®
liquid
AntifectNliquid
Meliseptol
®
Biguanide Incidin
Plus
Aldehyde KohrsolinExtra
Oxidizingagents Dismozon
®
pur
Phenol Helipur
®
Activchlor Trichlorol
®