EasyManua.ls Logo

LoveGo LG103 - User Manual

Default Icon
10 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
G3
Portable oxygen concentrator
User Manual
Lovego, Lets Go Along With Love
Table of contents
1. Introduction·················································································2
2. Important Safety Rules·································································3
3. Unpack your Lovego·····································································4
4. User Controls & System Status Indicators······································5
5. Quick Start···················································································6
6. How to operate·············································································7
7. How to operate with battery··························································9
8. How to operate with battery·························································10
9. Important Safety Rules·································································11
10. Maintenance··············································································13
11. Trouble shooting········································································14
12. Specification··············································································16
13. Warranty····················································································17
14. Online support···········································································18

Questions and Answers

  • S
    Shannon HillAug 8, 2025
    Why is my LoveGo Medical Equipment beeping with both green and red lights on?
    • N
      nicholasmcintyreAug 8, 2025
      The beeping alarm with solid green and red lights indicates a System HOT Shut Down. This means the concentrator's temperature is too high, causing oxygen production to stop. Ensure the bottom outlet vents are not blocked to allow proper airflow. If the problem continues, use a backup oxygen source and contact your equipment provider.
  • D
    David HicksAug 12, 2025
    What to do if my LoveGo LG103 alarm beeps but the LCD display is off?
    • N
      nbauerAug 12, 2025
      If the alarm is beeping but the LCD display isn't lit, the issue may be with the power supply. First, check the AC power supply connection to ensure it is properly connected. If you're using DC power, verify the battery connection and the connection at the cigarette lighter.
  • A
    annthomasAug 18, 2025
    Why is the yellow light on my LoveGo Medical Equipment solid for more than 5 minutes?
    • S
      Sue MoranAug 18, 2025
      A solid yellow light for more than 5 minutes indicates that the oxygen purity is low, meaning the concentrator is producing oxygen at a low level (less than 82%). Contact your equipment provider for assistance.
  • C
    Courtney PhillipsAug 22, 2025
    What does it mean if the red light on my LoveGo LG103 Medical Equipment stays on for more than 5 minutes?
    • L
      Linda KellerAug 22, 2025
      A solid red light for more than 5 minutes indicates low oxygen purity, specifically that the concentrator is producing oxygen at a low level (less than 72%). Contact your equipment provider.
  • J
    jamesmillerAug 28, 2025
    How to fix a 'no breath' alarm on LoveGo LG103 Medical Equipment?
    • A
      Adriana WeaverAug 28, 2025
      If you're experiencing a 'no breath' alarm, it could be due to a problem with the cannula or the use of a humidifier: * Check the cannula and its connection to the concentrator nozzle to ensure it's properly connected and not kinked or obstructed. * If you are using a humidifier, discontinue its use when on pulse flow.
  • S
    Sierra HoffmanSep 1, 2025
    What to do if my LoveGo LG103 Medical Equipment doesn't turn on when I press the power button?
    • S
      samantha91Sep 1, 2025
      If the concentrator does not power on when pressing the On/Off button, check the power switch located on the side of the machine to ensure it is turned on.

Summary

Important Safety Rules

Safety Symbols Explained

Explains the meaning and significance of various safety symbols used in the manual and on the device.

Quick Start

Initial Setup Steps

Outlines the basic steps required to connect and start the oxygen concentrator for the first time.

How to Operate Your Lovego POC

Operating Controls and Indicators

Describes the function of the device's controls, buttons, and status indicators during operation.

How to operate with battery and in the car

Continued Safety Precautions

Electrical Shock and Flammability Hazards

Details critical safety warnings regarding electrical shock, use near flammable materials, and oxygen risks.

Troubleshooting

Common Problems and Solutions

Lists common issues with the concentrator and their recommended solutions.

Alarm Indicators and Causes

Explains the meaning of various alarms and their potential causes.

LoveGo LG103 Specifications

General IconGeneral
BrandLoveGo
ModelLG103
CategoryMedical Equipment
LanguageEnglish

Related product manuals