What to do if my Lowenstein Medical prismaAQUA switches off?
- JJeremy HansenAug 2, 2025
If your Lowenstein Medical Humidifier keeps switching off, it's likely because the humidifier is empty. To resolve this, simply fill the humidifier with water.
What to do if my Lowenstein Medical prismaAQUA switches off?
If your Lowenstein Medical Humidifier keeps switching off, it's likely because the humidifier is empty. To resolve this, simply fill the humidifier with water.
Why is my Lowenstein Medical prismaAQUA Humidifier leaking?
If your Lowenstein Medical Humidifier is leaking, it could be due to several reasons: * A faulty heater rod seal. In this case you should replace the seal. * Incorrectly inserted humidifier insert. Make sure to insert the humidifier insert correctly. * A faulty humidifier insert, which means you'll need to replace it. * Cracks in the lower part of the humidifier, requiring replacement of the lower part.
How to fix Lowenstein Medical prismaAQUA that does not warm up?
If your Lowenstein Medical Humidifier isn't warming up, there are a couple of potential reasons: * The humidifier stage might be switched off. Ensure you set the humidifier stage correctly. * The humidifier itself could be faulty. In this case, you should have the humidifier repaired.
Heated Breathing Tube | Yes |
---|---|
Preheating | Yes |
Type | Heated Humidifier |
Power Consumption | 60 W |
Heating System | Integrated heated plate |
Humidification Modes | Manual Humidification |
Power Supply | 100 – 240 V AC, 50/60 Hz |
Instructions for filling the humidifier, including alternative methods.
Guide on how to connect the humidifier to the therapy device.
Steps for safely removing the humidifier from the therapy device.
Instructions on how to take the humidifier apart.
Steps for putting the humidifier back together.
Procedure for replacing the seal on the heater rod.
Defines the purpose and applications of the humidifier.
Explains how the humidifier works using the "pass-over" principle.
Lists conditions for which the humidifier is recommended.
Specifies conditions where the humidifier should not be used.
Information regarding known side effects of the humidifier.
Provides essential safety guidelines and warnings for humidifier use.
Identifies the individual parts of the humidifier using a diagram key.
Explains the meaning of symbols found on the device and its ID plate.
Details symbols on the device's identification plate.
Step-by-step guide on how to remove scale buildup from the humidifier.
Instructions for replacing the heater rod seal.
Information on whether the humidifier requires servicing.
Details the items included in the product package.
Information on ordering additional accessories and replacement parts.