Operating Manual Smart Weather Sensor
102 G. Lufft Mess- und Regeltechnik GmbH, Fellbach, Germany
19.6.4.10 Buffer Assignment Basic Data Set Power Saving Mode 2 (all Models)
Device configured for measurement in metric units:
Measurement Value
UMB
Channel
Min
Max
Unit
Buffer ‘0’
Air Temperature (act)
100
-50.0
60.0
°C
Rel. Humidity (act)
200
0.0
100.0
%
Amount of Precip. Difference
625
0.00
100000.00
mm
Rel. Air Pressure(act)
305
300.0
1200.0
hPa
Wind Speed (act)
400
0.0
75.0
m/s
Buffer ‘1’
Wind Direction(act)
500
0.0
359.9
°
Global Radiation
900
0.0
2000.0 *)
W/m
2
Leaf Wetness State (act)
711
0 / 1
External Temperature
101
-20.0
80.0
°C
Buffer ‘2’
Amount of Precip. Absolute
620
0.0
100000.0
mm
Dew point (act)
110
-50.0
60.0
°C
Abs. Humidity (act)
205
0.0
1000.0
g/m³
Mixing Ratio(act)
210
0.0
1000.0
g/kg
Abs. Air Pressure(act)
300
300.0
1200.0
hPa
Buffer ‘3’
Wet Bulb Temperature (act)
114
-50.0
60.0
°C
Specific Enthalpy (act)
215
-100.0
1000.0
kJ/kg
Air Density
310
0.000
3.000
kg/m3
Compass (act)
510
0.0
359.0
°
*) Until device version .49 / .212 / Firmware v5.6: 1400.0 W/m
2
Example: Request Buffer ‘0’
0D0!
0+13.5+85.7+0.2+1017.0+1.8
Air temperature 13.5°C, rel. humidity 85.7%, precipitation 0.2°C, rel. air pressure 1017.0hPa, wind speed1.8m/s