MOTOR ASSEMBLY









Remove motor slide caps
motor
slide cap
Remove motor slide caps
1. Remove the slide caps from both ends of the motor as shown above. Plungers should be recessed
in motor cavities; if not, plug the power cord into any grounded outlet and operate “down”
function on the pendant hand control, then unplug the power cord.
Info: Motor comes from factory with pendant hand control plugged into motor.
2. Install 9V backup battery (not provided) and slide into storage area.
Install Motors
Install motor
cam
tap here
foot
head
3. Hold the bed securely, positioned with motor cams up, as shown above.
4. Position the motor over both cams.
5. Tap either end of the motor housing (shown below) with enough force to snap the motor
assembly into position on the cam.
6. Repeat step 5 for other end of motor.
motor
slide cap
Install motor slide caps
pressure