10
Technical specifications
Power supply
AC input voltage 100 to 240 VAC
AC input current Max 1 A @100–240 VAC
AC frequency 50 to 60 Hz
DC input voltage 5 VDC, 0�5 A max
Power consumption 100 W max
Fuses Quick-Acting F4A H, 250 VAC (H = High Breaking
Capacity (Ceramic Tube))
Operation and Maintenance
Interfaces 1 RJ45 port LAN 1000BASE-T
USB 3�0
Dimensions (H × W × D) 505 mm × 430 mm × 400 mm
Weight 25�5 kg
Environmental specications For indoor use only, +5ºC to +40ºC (+41ºF to +104ºF)
Altitude up to 2000 m
Relative humidity range: 5–95%
Pollution Degree 2
Mains supply tolerance (%) or absolute
mains supply values (voltage uctuations
of the nominal voltage)
+/-10%
Transient overvoltages Overvoltage Category II
Surge Level 3 (CM 12 ohm) & 4 (DM 2 ohm): 2 kV– 1�2/50
µs
IP class IP20
Key features
Read/calculation speed ELISpot < 2 min/plate
FluoroSpot 1 color < 5 min/plate
FluoroSpot 2 colors < 7 min/plate
FluoroSpot 3 colors < 10 min/plate
FluoroSpot 4 colors < 13 min/plate
Resolution (H × W) 2048 × 2048 pixels
Approved plate types 96-well lter plates with 0.45 µm PVDF membrane:
IPFL, MSIP, or MAIPSWU10 (with an adapter)
The Mabtech IRIS
TM
FluoroSpot/ELISpot reader must be connected to a computer
equipped with the Mabtech Apex
TM
-software�