Setup and installation
Warning! Follow the instructions in this manual when you lift or move the
reader� Risk of injury to person or damage to the system�
Warning! Do not use or keep ammable liquids near the system. Risk of
re.
Caution! The reader is heavy� Place it on a solid table� Use two people to lift
or move the reader
Caution! Keep other equipment more than 10 cm away from the reader
Risk of malfunction caused by overheating�
Note: Keep the packaging in case the system needs to be sent back to
Mabtech for service
1. Lift the Mabtech IRIS
TM
FluoroSpot/
ELISpot reader out of the box by the
handles
2. Place the Mabtech IRIS
TM
FluoroSpot/
ELISpot reader upright on a solid table
Remove the inner packaging� Make
sure that you have received all
components specied in the section “In
this package” on page 8�