14
High voltage
application is set to
run on a very low
current
DISABLE No Load alarm on the
Display Screen
Error Messages
Err1; power
supply stops;
alarm sounds.
Current exceeds
setting
- Check that the correct
running buffer and buffer
concentration have been
used
- Check that cables are
connected properly
- Switch power on / off and
then restart application
- If the power supply cannot be
restarted, disconnect the
power supply from the mains
outlet and contact Technical
Service
Err2; power
supply stops;
alarm sounds.
Voltage exceeds
setting
- Check that the correct
running buffer and buffer
concentration have been
used
- Check that cables are
connected properly
- Switch power on / off and
then restart application
- If the power supply cannot be
restarted, disconnect the
power supply from the mains
outlet and contact Technical
Service