parameters E1 and E2
2 – as in setting 1, but disabling with the
stirrer's working button causes its operation
only during the operation of the unit
3 – as in setting 1, but the button that disables
the stirrer operation is inactive
4 – continuous operation, regardless of the
operation of the unit and times set in the
parameters E1 and E2
standstill time of the stirrer
the possibility of setting the time from 1 min.
to 60 min., every 1 min.
working time of the stirrer
the possibility of setting the time from 1 min.
to 60 min., every 1 min.
manual stirrer operation
mode
0 – continuous work
>0 – stirrer operation for a specified time in the
range of from 1 min. to 60 min., every 1 minute
delay of registering of the
max. / min. values
of temperature and delay
of activation of the
temperature alarm
the possibility of setting the time from 0 h to
24 h, every 0,1 h
lower temperature of the
alarm
the possibility of setting the temperature from
-40
O
C to +120
O
C, every 1.0
O
C
upper temperature of the
alarm
the possibility of setting the temperature from
-40
O
C to +120
O
C, every 1.0
O
C
visual and sound alarm
of exceeding the set
temperature range
(parameters A1 and A2)
scaling the control sensor**
setting the operation of the
limit switch of opening the
tank’s flap
0 – closing the limit switch: raising the flap
1 – opening the limit switch: raising the flap
delay of switching on the unit
and the stirrer after
connecting
the regulator to the mains or
when power supply decay
occurs
from 1 s. to 999 s., every 1 s.
display status when the
control is switched off
0 – the dot is displayed
1 – the currently measured temperature is
displayed
return to factory settings
0
1 – after setting "1" and turning off the
regulator from the power supply, switching on
again restores the factory settings and the
parameter value goes back to "0"
** set at the manufacturing stage (do not change without obvious need)