manual stirrer operation
mode
CP
0.0 continuous work
>0.0 stirrer operation for a specified time in
the range of from 1 min. to 60 min., every 1
min.
delay of registering of the
maximum / minimum
values of temperature and
delay of activation of the
temperature alarm
AA
setting the time from 0 h to 4 h, every 0,1 h
lower temperature of the
alarm
A1
setting the temperature from -40
O
C to
+120
O
C, every 1.0
O
C
upper temperature of the
alarm
A2
setting the temperature from -40
O
C to
+120
O
C, every 1.0
O
C
sound of the alarm
HE
0 active
1 inactive
visual and sound alarm
of exceeding the set
temperature range
(parameters A1 and A2)
HA
0 active
1 inactive
scaling the control
sensor**
CU
every 0,2
O
C
setting the operation
of the limit switch
of opening the tank’s flap
UU
0 closing the limit switch: raising the flap
1 opening the limit switch: raising the flap
delay of switching on the
chiller (unit) and the stirrer
after connecting the
regulator to the mains
or when power supply
decay occurs
dE
setting the time from 1 sec. to 999 sec., every
1 sec.
display status when
the control is switched off
d1
0 the dot is displayed
1 the currently measured temperature is
displayed
return to factory settings
FA
0 no return to the factory settings
1 after setting parameter to 1 and turning
off the regulator from the power supply,
switching it on again restores the factory
settings and the parameter value
automatically goes back to 0
leaving the settings
EE
** set at the manufacturing stage (do not change without obvious need)