STERILMATIC
ANALOG ELECTRIC STERILIZER
INSTALLATION - OPERATION - MAINTENANCE
Telephone: (802) 658-6600 Fax: (802) 864-0183
www.marketforge.com PN 14-0411 Rev D (2/16)
© 2015 - Market Forge
MODELS
STM-E
STM-EL
STM-EX
STM-ELX

Questions and Answers

  • P
    pauledwardsAug 12, 2025
    What to do if my Market Forge Industries STM-EL Laboratory Equipment sterilizer won't turn on?
    • V
      Vincent ReynoldsAug 12, 2025
      If your sterilizer isn't operating and there's no pressure build-up, here's what you can check: * First, ensure it's installed correctly by checking the wire diagram. * Second, replace the fuse, and if it blows again, verify the electrical supply is 60 amps. * Third, the contactor may be burned out, in which case it should be replaced. * Finally, inspect all wiring for defects and repair or replace as necessary.
  • R
    rebecca86Aug 17, 2025
    Why my Market Forge Industries STM-EL sterilizer is not building pressure?
    • W
      watkinsericaAug 18, 2025
      If your sterilizer operates but fails to reach 15.5 PSI pressure, consider the following: 1. The current may not be heating all elements. Remove the lower front panel to check if the heating elements are working. 2. The steam trap may not be closing. Replace the thermometer within the steam trap. 3. The exhaust valve may not be holding pressure. Check and readjust the temperature control if it's incorrectly adjusted. 4. Steam could be leaking around the door. Inspect the gasket for wear or adjust the door. 5. The safety valve might be blowing off prematurely. Replace the safety valve.
  • A
    Amanda GutierrezAug 25, 2025
    How to fix Market Forge Industries STM-EL Laboratory Equipment heating elements staying on?
    • J
      Jerry MillerAug 25, 2025
      If one or both sides of the cast-in heating elements remain on when the timer is off: * The contactors of the temperature control switch may be stuck closed, requiring switch replacement. * Alternatively, the temperature control might not be properly calibrated, so recalibration is necessary.
  • T
    Taylor DuncanAug 30, 2025
    What to do if Market Forge Industries STM-EL Laboratory Equipment releases pressure too early?
    • K
      Kylie EvansAug 30, 2025
      If your unit releases pressure before the cycle has completed, the low water cut-off may have functioned prematurely. If so, replace the low water cut-off.
  • N
    Natalie PageSep 4, 2025
    Why is my Market Forge Industries STM-EL Laboratory Equipment heating element cutting out too soon?
    • L
      Lori BurnsSep 5, 2025
      If the heating element cuts out before reaching 15 lbs of pressure, the pressure might be cutting off too soon at the temperature control. Readjust the range of operation and calibration of the temperature control.

Market Forge Industries STM-EL Specifications

General IconGeneral
ModelSTM-EL
CategoryLaboratory Equipment
ManufacturerMarket Forge Industries
TypeSteam Sterilizer
Chamber MaterialStainless Steel
Door TypeManual