STERILMATIC
ANALOG ELECTRIC STERILIZER
INSTALLATION - OPERATION - MAINTENANCE
Telephone: (802) 658-6600 Fax: (802) 864-0183
www.marketforge.com PN 14-0411 Rev D (2/16)
© 2015 - Market Forge
MODELS
STM-E
STM-EL
STM-EX
STM-ELX

Questions and Answers

  • M
    Michael HoldenAug 11, 2025
    What to do if my Market Forge Industries STM-EX sterilizer won't turn on?
    • J
      joshuareynoldsAug 12, 2025
      If your sterilizer isn't operating and there's no pressure build-up, here's what you can check: * First, ensure it's installed correctly by checking the wire diagram. * Second, replace the fuse, and if it blows again, verify the electrical supply is 60 amps. * Third, the contactor may be burned out, in which case it should be replaced. * Finally, inspect all wiring for defects and repair or replace as necessary.
  • C
    Clarence McdonaldAug 17, 2025
    Why my Market Forge Industries Laboratory Equipment sterilizer is not building pressure?
    • P
      padamsAug 18, 2025
      If your sterilizer operates but fails to reach 15.5 PSI pressure, consider the following: 1. The current may not be heating all elements. Remove the lower front panel to check if the heating elements are working. 2. The steam trap may not be closing. Replace the thermometer within the steam trap. 3. The exhaust valve may not be holding pressure. Check and readjust the temperature control if it's incorrectly adjusted. 4. Steam could be leaking around the door. Inspect the gasket for wear or adjust the door. 5. The safety valve might be blowing off prematurely. Replace the safety valve.
  • R
    Ryan HollandAug 25, 2025
    How to fix Market Forge Industries STM-EX Laboratory Equipment heating elements staying on?
    • S
      stewartjulianAug 25, 2025
      If one or both sides of the cast-in heating elements remain on when the timer is off: * The contactors of the temperature control switch may be stuck closed, requiring switch replacement. * Alternatively, the temperature control might not be properly calibrated, so recalibration is necessary.
  • A
    alexis47Aug 29, 2025
    What to do if Market Forge Industries STM-EX Laboratory Equipment releases pressure too early?
    • J
      johnsonjenniferAug 30, 2025
      If your unit releases pressure before the cycle has completed, the low water cut-off may have functioned prematurely. If so, replace the low water cut-off.
  • T
    Timothy EvansSep 5, 2025
    Why is my Market Forge Industries Laboratory Equipment heating element cutting out too soon?
    • A
      Alan NunezSep 5, 2025
      If the heating element cuts out before reaching 15 lbs of pressure, the pressure might be cutting off too soon at the temperature control. Readjust the range of operation and calibration of the temperature control.