4
1.2 Specifications
Parameter Range Accuracy Resolution
Conductivity
0-1.000 µS/cm +0.01 µS/cm 0.001 µS/cm
0-10.00 µS/cm +0.1 µS/cm 0.01 µS/cm
0-100 µS/cm +1.0 µS/cm 0.1 µS/cm
Temperature 0-100
o
C +0.1
o
C 0.01
o
C
Power: 97-240 VAC, 50/60 Hz, 24Vdc, fuse protected
Recorder Output: Four isolated, scaleable 0-20 mA DC analog & RS232C serial ASCII digital output
Alarm Output: Four solid-state relays, 5 amp, 250 V
Weight: 2 lbs (0.9 kg) including one sensor with 10 feet (3 meters)
Construction: Readout Module Case: ABS plastic, NEMA 12 (IP65)
Sensor: Type 316 Stainless Steel and Teflon
Cabling: 8-conductor, 24 gauge, PVC jacket, shielded, 0.25 inch (0.63 cm)
Dimensions: ¼ DIN case, 6.51 x 3.78 x 3.78 (165.5 x 96 x 96 mm) LWH
Installation: Panel mount (mounting hardware included). Recommend mounting sensor
vertically with inlet at the bottom. Position of sensor cabling not a factor in
mounting.
Flow rate: 45-250 ml/min