EasyManua.ls Logo

Martel MC1210 - Page 42

Martel MC1210
54 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
42
CPRT_COEFA?
Returns the number which was entered for the first coefficient of the polynomial used in the
custom RTD. Using the example above CPRT_COEFA? Would return:
3.908000E-03
CPRT_COEFB
This command is used for entering a custom RTD into the calibrator. The numeric value
entered after the command will be set as the second coefficient of the polynomial used by the
custom RTD.
For example:
CPRT_COEFB -5.8019E-07 enters -5.8019e-7 as coefficient B.
CPRT_COEFB?
Returns the number, which was entered for the first coefficient of the polynomial used in the
custom RTD. Using the example above, CPRT_COEFB? Would return:
-5.801900E-07
CPRT_COEFC
This command is used for entering a custom RTD into the calibrator. The numeric value
entered after the command will be set as the first coefficient of the polynomial used by the
custom RTD.
For example:
CPRT_COEFC -5.8019E-12 enters -5.8019e-12 as coefficient A.
CPRT_COEFC?
Returns the number which was entered for the first coefficient of the polynomial used in the
custom RTD. Using the example above CPRT_COEFC? Would return:
-5.801900E-12
CPRT_MIN_T
Sets the minimum temperature of the custom RTD range. The temperature value must be
entered with a degrees label, CEL for Celsius and FAR for Fahrenheit.
For example:
CPRT_MIN_T -260 CEL enters -260°C as the minimum temperature.
CPRT_MIN_T?
Returns the value entered for minimum temperature in the range for a custom RTD. Note that
the Calibrator always returns numbers in scientific notation. The above example would return:
-2.600000E+02, CEL
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com