73
8995B-eIFU-1115
5 10 15 20 25 30 35 40
5
10
15
20
25
30
35
40
0 < Number of Samples < 10
10 Number of Samples < 100
100 Number of Samples
Control (Respiration per Minute)
Capnograph (Respiration per Minute)
Healthy Adult Subjects: Capnograph vs. Control (manual count)
5 10 15 20 25
−30
−20
−10
0
10
20
30
40
Bias=0.22, STDEV=1.62
(Capnograph + Control)/2 (Respiration per Minute)
Capnograph − Control (Respiration per Minute)
0 < Number of Samples < 10
10 Number of Samples < 100
100 Number of Samples < 1000
1000 Number of Samples
Healthy Adult Subjects: Capnograph vs. Control (manual count)
Respiration Per Minute (RPM) Difference when Compared to
Control for Healthy Adult Subjects
Coverage
RAM Capnograph
+ 1 RPM 74.1% 70.0%
+ 2 RPM 90.4% 88.4%
+ 3 RPM 95.6% 94.9%