MS8218 True RMS Multimeter
Technological Specifications
AC Current/DC Current + AC Current
Accuracy Voltage Drop
Range Resolution
40Hz
1KHz 1KHz
10KHz
10KHz
20KHz
500
μ
A 0.01
μ
A
±
(0.75%+20)
±
(1%+20)
±
(2%+20)
5000
μ
A 0.1
μ
A
±
(0.75%+10)
±
(1%+10)
±
(2%+10)
102µV/µA
50mA
1
μ
A
±
(0.75%+20)
±
(1%+20)
±
(2%+20)
500mA
10
μ
A
±
(0.75%+10)
±
(1%+10)
±
(2%+10)
1.5mV/mA
5A 0.1mA
±
(0.75%+20)
±
(1.5%+20)
±
(5%+20)
10A 1mA
±
(1.0%+10)
±
(1.5%+10)
Unspecified
30mV/A
Notes: above accuracies can be guaranteed within 10%-100% of the full range
DC Current
Range Resolution Accuracy Voltage Drop
500
μ
A 0.01
μ
A
±
(0.15%+15)
5000
μ
A 0.1
μ
A
±
(0.15%+10)
102µV/µA
50mA
1
μ
A
±
(0.15%+10)
500mA
10
μ
A
±
(0.15%+10)
1.5mV/mA
5A 0.1mA
±
(0.5%+10)
10A 1mA
±
(0.5%+10)
30mV/A
Notes: above accuracies can be guaranteed within`the full range