Portable Sampling Drive Operating ManualMasterflex
Section 7 Specifications
7-1
Operating Speed 70 to 400 rpm
Operating Voltage/Frequency:
Model 07571-02 115V AC nominal 50/60 Hz (100–130) V AC @ 120mA AC
230V AC nominal 50/60 Hz (200–260) V AC @ 60mA AC
12V DC nominal (11–16)V DC @2.4A DC
Number of Pump Heads: 1
Torque Load: 50 oz-in (0.35 N•m), maximum
Direction of Rotation: Clockwise and Counterclockwise
Operating Temperature: 32°F to 122°F (0° to 50°C)
Storage Temperature: −4°F to 149°F (−20°C to 65°C)
Humidity (Non-condensing): 10% to 90%
Altitude 6600 ft. (2000 m)
Dimensions (L x W x H): 17.9 in x 12.1 in x 9.6 in
(38 cm x 31 x 24 cm)
Weight: 17.8 pounds (8.07 kg)
Chemical Resistance: High-density polyethylene, clear anodized aluminum,
stainless steel and polyester. All materials withstand
standard solvents.
Enclosure Rating: IP54 per IEC529
Pollution Degree: Degree 2 per IEC664
Installation Category: Category II per IEC664 (local level—appliances, portable
equipment, etc.)
Compliance:
07571-02 115V AC Power Supply AC/DC Converter is UL listed and
CSA approved. Regulatory agency specifications not
applicable to the balance of the unit due to low voltage.
CE Mark:
EN809 (EU Machinery Directive)
EU EMC Directive
(EN61326-1)