866.4.Maxtec www.maxtec.com8
7.2 Effect of Interferent Gases and Vapors
Gas Volume % Dry Interference in O
2
Nitrous Oxide 75% < 2%
Halothane 5% < 2%
Isoflurance 5% < 2%
Enflurane 5% < 2%
Sevoflurane 6% < 2%
Desflurane 15% < 2%
Carbon Dioxide 10% < 2%
Helium 70% < 2%
8.0 HANDI+ SPARE PARTS AND ACCESSORIES
8.1 Standard Replacement Parts and Accessories
Part Number Item
R218M12 Operation Manual, Handi+
R218M12-001 Operation Manual, Handi+
International
RP16P02 "T" Adapter
R110P10-001 Flow Diverter Fitting
R207P17 Barbed Adapter
R100P92-002 Clear Tubing
R213P92 Lanyard
RP16P10 MaxConnect “T” Adapter
R218P09 Handi+ Cover
2.45 in
0.90 in
1.44 in
2.91 in