Do you have a question about the Maxtec UltraMax O2 and is the answer not in the manual?
Explains the analyzer's design, features, and operational benefits.
Defines the intended use of the UltraMax 02 for servicing oxygen concentrators.
Identifies and describes the main parts and display elements of the device.
Step-by-step guide on how to measure oxygen concentration, flow, and pressure.
Instructions on how to toggle between PSI and kPa for pressure readings.
Procedure to verify the accuracy of the UltraMax 02 using a pure oxygen source.
How temperature changes can affect measurement accuracy.
Impact of humidity on the device's accuracy and potential damage.
How non-standard gas mixtures can lead to incorrect readings.
How low flow rates can alter oxygen concentrator output and affect measurements.
Instructions for replacing the device's batteries.
Guidelines for cleaning the exterior and general device care.
Technical details and measurement ranges for oxygen, flow, and pressure.
Lists included items and optional accessories for the UltraMax 02.
Details on electromagnetic emissions and their compliance standards.
Test levels and environments for immunity against ESD, transients, surges, and dips.
Recommended separation distances for RF equipment based on output power.
| Measurement Range | 0-100% |
|---|---|
| Resolution | 0.1% |
| Sensor Type | Galvanic fuel cell |
| Accuracy | ±1% |
| Response Time | <15 seconds |
| Operating Temperature | 15°C to 40°C (59°F to 104°F) |
| Storage Temperature | -15°C to 50°C (5°F - 122°F) |
| Operating Humidity | 0-95% RH, non-condensing |
| Power Supply | 2 AA alkaline batteries |
| Battery Life | > 5, 000 hours continuous use |