6 | PRM-8000 Quick Start Guide
Sub-Menu
Selection:DEFAULT
Description
Values - Press
to change/select
Utilities
Reset Counters Resets Current, Minimum, Maxi-
mum, Average, Total Count and
Elapsed Time registers
Hardware Reset Resets device and clears all reg-
isters including Time and Date
Power Down Turns power off and shuts down
all radiation monitoring functions
Load Defaults Loads factory calibration NOTE: This function will
erase existing calibration
data. Use with care.
Display Bat/DC Displays battery and external DC
input voltages
Calibrate Enters the calibration routine.
Two calibrated dose rates are re-
quired - 20mR/hr and 200mR/hr
NOTE: Calibration should
be performed by a certi-
fi ed lab with a certifi ed
Cs137 source.
Ext Audio/Data
Off Provisions the external audio
and data jack for use with either
headphones or the optional USB
interface cable. The Pulse Out
setting can be used to interface
the PRM-8000 to radiation
graphing software or to other
electronic equipment.
Off, Data, Audio:Low, Au-
dio: Medium, Audio: High,
and Pulse Out
Data Logging
Every: Minute Sets how often a measurement
is stored in the data log memory
Minute or Hour
Start or Stop Starts Logging, or Stops Logging
Status Shows the number of Records
and Memory remaining during
logging and after logging has
been stopped.
Send Data Transmits log data in comma
separated value format at 57,600
baud (requires data cable)
NOTE: This command is
visible only when logging
is stopped.
Language ゲンゴ
English Used to select either English or
Japanese
English, Japanese