Operating Manual
Operation instructions Page 3-10
3.4 Connecting the System
CAUTION
Risk of damage!
► Exact pressure setting required.
Overpressure will damage the system - low pressure will
cause malfunction.
!
NOTE!
Large systems may require having the working gas split into Working
Gas and Control Gas connections.
3.3.5 Equipment for Vacuum Pumps
Disposal Piping for Vacuum Pump Waste Gas:
Exhaust Outlet for Waste Regeneration Gas:
Use
Connecting the
Regeneration Gas Exhaust system connection with
vacuum pump
Material
Optional (length as required):
either Ø 9 mm reinforced hose, 3 mm wall thickness and adapter,
or Ø 10 mm copper pipe or Ø 10 mm stainless steel pipe
Connection
Ø 9 mm hose nozzle or Ø 10 mm Swagelok fitting.
Type
Use
Connecting the vacuum pump exhaust (oil mist and waste gas) with the
customer’s waste gas disposal facility (depressurized exhaust
outlet).
Material
Optional (length as required):
eith
er: Ø 25 mm reinforced hose and Ø 25 mm hose nozzle
or: Ø 25 mm copper pipe as well as flange and clamp
or: Ø 25 mm stainless steel pipe as well as flange and clamp.