BioStamp nPoint User Manual
Guidance and Manufacturers Declaration – Electromagnetic Immunity
The BioStamp nPoint system is intended for use in the electromagnetic
environment specied below. The customer or the user of the BioStamp
nPoint system should assure that it is used in such an environment.
Immunity Test IEC 60601 Test
Level
Compliance
Level
Electromagnetic Envi-
ronment - Guidance
Electrostatic
Discharge (ESD)
EN61000-4-2
+/-15 kV Air
Discharge
+/-8 kV Contact
Discharge, VCP,
HCP
+/-15 kV
+/-8 kV
Floor should be wood,
concrete or
ceramic tile. If oors
are covered with
synthetic material,
the relative humidity
should be at least 30%.
Radiated
Electromagnetic
Fields
EN61000-4-3
10 V/m, 80-2700
MHz at 80%, 1
kHz AM
Modulation
10 V/m
Portable RF
communications
equipment (including
peripherals such as
antenna cables and
external antennas),
microwave ovens,
cordless phones, Wi-Fi
devices should be used
no closer than 30 cm
(12 inches) to any part
of BioStamp nPoint
system, including
cables specied by the
manufacturer,
otherwise,
degradation of the
performance of this
equipment could
result.
Radiated
Electromagnetic
and Proximity
Fields
EN610004-3
9-28 V/m, RF
Wireless
Communication
Fields on Spot
Frequencies from
Table 9 at 50%,
Square Wave
Modulation
9-28 V/m
Conducted
EN61000-4-6
6 Vrms on ISM
and Amateur
Bands
3 Vrms, 0.15-80
MHz, AC Mains
3 Vrms, 0.15-80
MHz, SIP/SOP
Ports
6 Vrms
3 Vrms
3 Vrms