EasyManua.ls Logo

Mecmesin AFG Series - % DROP Feature; 1 st Peak Detection; 1 st Peak Tension;Compression

Mecmesin AFG Series
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
page 19
The display will show:-
STAND OFF/ON -Indicates status of stand control function.
REVERSE - Reverses the stand direction of travel at
sample break (BREAK) or load-limit value
(LIMIT). The test stand will reverse back to
the start position as defined by the
physical Microswitch.
STOP - Stops the stand at sample break (BREAK)
or load-limit value (LIMIT). The test stand
does not return to the start position.
CYCLE - Cycles a suitable test stand between load
limits (UPPER, LOWER) for a set number of
times (CYCLE).
Select the desired function using UP and DOWN keys and
press the ENTER key to select.
Select UP or DOWN to tell the gauge which direction the
stand will begin to move before the load-limit is reached.
BREAK - Sets the gauge to reverse at sample break. Set %
break using UP/DOWN arrow keys. Press ENTER to select.
SET % of load cell capacity to indicate the value by which the
load must fall to determine a break. Use a higher percentage
for ‘noisy’ samples where the load may fluctuate before the
sample finally breaks.
Press the ENTER key to confirm selection and return to stand
sub- menu 1.
LIMIT - Sets the load-limit value to trigger the stand reverse
function. Press ENTER key to select.
SET load-limit using UP and DOWN keys. (UNITS key
changes the units of measurement for load-limit value). Press
ENTER key to confirm selection and return to stand sub-menu.
Negative load limits indicate compression.
STAND sub-menu 1
The cycle function only refers to
the VersaTest (Quantrol TC2)
motorised test stand
REVERSE sub-menu 1
REVERSE sub-menu 2
Break sub-menu 1
Limit sub-menu 1

Other manuals for Mecmesin AFG Series