page 19
Common Specifications
Speed resolution ±1.0mm/min (±0.04in/min)
Displacement resolution 0.01mm or 0.0004 inches
Operating modes Manual, Single Cycle, Continuous Cycle
Up and Down control By membrane switch
Digital display of speed/displacement On LCD screen
Direction of travel indication on stand On LCD screen
Reverse on alarm point Yes, with appropriate force gauge and cable
Reverse on sample break Yes, with appropriate force gauge and cable
Limit switch repeatability**** <0.5mm (0.02”)
Limit switch over-run at top speed <2mm (0.08”)
Operating temperature 10C to 35C (50F - 95F)
Humidity range Normal industry and laboratory conditions
Loadholding No
Graphical representation***** via Emperor™ Lite Software
Output of test results to via Emperor™ Lite Softwar
e
PC/Printer/Datalogger*****
Options (Available upon request)
Column gaiter
Extended throat depth
Safety guard
Horizontal operation
**** at given speed (i.e. not necessarily over whole range)
***** when used with a force gauge
Mecmesin reserves the right to alter equipment specifications without prior notice.
E&OE