Manual in the relative chapter for instructions of
cleaning and disinfection.
• Do not have the oximeter immerged in liquid.
When it needs cleaning, please wipe its surface
with medical alcohol by soft material. Do not spray
any liquid on the device directly.
• When cleaning the device with water, the tempe-
rature should be lower than 60°C.
• As to the ngers which are too thin or too cold, it
would probably affect the normal measure of the
patients‘ SpO
2 and pulse rate, please clip the thick
nger such as thumb and middle nger deeply
enough into the probe.
• Do not use the device on infant or neonatal
patients.
• The product is suitable for children above four
years old and adult (Weight should be between 15
kg to 110 kg).
• The device may not work for all patients. If you are
unable to achieve stable readings, discontinue use.
• The update period of data is less than 5 seconds,
which is changeable according to different indivi-
dual pulse rate.
• Please read the measured value when the wave-
form on screen is equably and steady-going. This