Medlab medizinische Diagnosegeräte GmbH EG01010 User Manual
Version 1.06
10
Pacemaker pulse rejection, section 201.12.1.101.13
Without over/undershoot:
a) For single (ventricular-only) pacemaker pulses alone, with 0.1 and 2.0 ms pulse-widths and ± 2 mV
and ± 700 mV pulse-amplitudes, the EG01010 correctly displays heart rate as zero bpm (Asystole).
b) For single (ventricular-only) pacemaker pulses with normally paced QRS-T, with 0.1 and 2.0 ms pulse-
widths and ± 2 mV and ± 700 mV pulse-amplitudes, the EG01010 correctly displays heart rate of the
QRS-T rhythm (60 bpm for the specified test waveform).
c) For single (ventricular-only) pacemaker pulses with ineffectively paced QRS pattern, with 0.1 and 2.0
ms pulse-widths and ± 2 mV and ± 700 mV pulse-amplitudes, the EG01010 correctly displays heart rate
of the underlying QRS-T rhythm (30 bpm).
d) For atrial/ventricular pacemaker pulses alone, with 0.1 and 2.0 ms pulse-widths and ± 2 mV and
± 700 mV pulse-amplitudes, the EG01010 correctly displays heart rate of zero bpm (Asystole).
e) For atrial/ventricular pacemaker pulses with normally paced QRS-T, with 0.1 and 2.0 ms pulse-widths
and ± 2 mV and ± 700 mV pulse-amplitudes, the EG01010 correctly displays heart rate of the QRS-T
rhythm (60 bpm).
f) For atrial/ventricular pacemaker pulses with ineffectively paced QRS pattern, with 0.1 and 2.0 ms pulse-
widths and ± 2 mV and ± 700 mV pulse-amplitudes, the EG01010 correctly displays heart rate of the
underlying QRS-T rhythm (30 bpm).
With over/undershoot:
a) For single (ventricular-only) pacemaker pulses alone, with 0.1 and 2.0 ms pulse-widths and ± 2 mV
and ± 700 mV pulse-amplitudes, the EG01010 correctly displays heart rate of zero bpm (Asystole).
b) For single (ventricular-only) pacemaker pulses with normally paced QRS-T, with 0.1 and 2.0 ms pulse-
widths and ± 2 mV and ± 700 mV pulse-amplitudes, the EG01010 correctly displays heart rate of the
QRS-T rhythm (60 bpm).
c) For single (ventricular-only) pacemaker pulses with ineffectively paced QRS pattern, with 0.1 and 2.0
ms pulse-widths and ± 2 mV and ± 700 mV pulse-amplitudes, the EG01010 correctly displays heart rate
of the underlying QRS-T rhythm (30 bpm).
d) For atrial/ventricular pacemaker pulses alone, with 0.1 and 2.0 ms pulse-widths and ± 2 mV and
± 700 mV pulse-amplitudes, the EG01010 correctly displays heart rate of zero bpm (Asystole).