Medlab medizinische Diagnosegeräte GmbH EG01010 User Manual
Version 1.06
14
14
Set delay of the pulse trigger signal (active high, 33 ms duration):
„T“ Parameter: „0“ , „1“ , „2“ , or „9“
„0“ Delay of the pulse trigger signal 15 ms (reset value)
„1“ Delay of the pulse trigger signal 50 ms
„2“ Delay of the pulse trigger signal 100 ms
„9“ The signal triggers in the middle between R waves
Pacemaker detection on or off:
„P“ Parameter: „0“ , „1“
„0“ Pacemaker detection off
„1“ Pacemaker detection on (reset value)