Amp: „00“ Amplification stage 1 (lowest)
„01“ Amplification stage 2
„10“ Amplification stage 3
„11“ Amplification stage 4 (highest)
Filter1: „0“ EMG filter off
„1“ EMG filter on
Filter2: „00“ 50/60 Hz filter off
„01“ 50 Hz filter on
„10“ 60 Hz filter on
Status, Byte 6: N 0 Adult mode
1 Neonatal mode
S0..3 0000 Normal operation
S0..3 0001 Normal operation, pacemaker detected
S0..3 0100 Initializing
S0..3 0101 Searching for electrodes
S0..3 1000 Simulated output
S0..3 1010 Selftest error
Rest: Not used, but reserved
„Selftest error“ (10) errors are fatal. The module will never go into measuring state.
Status message 8 will be transmitted as long as simulated data is transmitted.