DC 00573 TSA 2 Quick Set Up and Installation Overview Rev.1 Date of issue: 10.06. 2019 Page 7 of 9
Always allow 15-20 seconds for trapped air in the system to be pumped towards the air vent screw
between pushing the syringe plunger.
8. Disconnect the syringe from the system and keep it for future use.
9. Tight the vent screw back to the system. ( 3 turns clockwise)
10. Important notes:
10.1. It is recommended to fill the system before it reaches the lower indicator (most left of the four LED lights).
10.2. When the coolant level is too low (coolant level empty) safety precautions are activated and the system will
not allow work to continue before filling the system.
System Requirements: 1 GHZ CPU frequency. Windows 7,8 or 10. 2GB RAM.
see section 4.1 in TSA 2 Operation Manual.
Note: Software installation can be performed only by Windows system administrators. Make sure
you have the appropriate administrator privileges on your computer. Or have your system
administrator to install the software for you.