PATHWAY
Pain & Sensory Evaluation
System
Technical Reference
Manual

Table of Contents

Questions and Answers

  • K
    Katelyn GarciaSep 18, 2025
    What to do if Sensor Heater\TEC missing calibration table entries in range 10 - 48 on Medoc PATHWAY Medical Equipment?
    • M
      Marcia SolisSep 18, 2025
      If the sensor calibration cannot be completed because there are missing points in the required range, complete the missing entries and continue.
  • M
    melissawillisAug 25, 2025
    What does it mean if the emergency button was pressed on Medoc PATHWAY Medical Equipment?
    • C
      Christopher BryantAug 25, 2025
      If the system detected that the emergency button has been pressed, or the hardware safety mechanism has been triggered: 1. Release the emergency button and/or the Emergency button on the extended cable. 2. Run a system Self-Test. * If the Self-Test completes without any further errors the system can return to normal operation. * If the Self-Test cannot be complete, restart the system and retry. If the Self-Test continues to fail, there is no other solution.
  • W
    William MataAug 25, 2025
    What to do if Medoc Medical Equipment Pathway is in Safe Mode?
    • B
      Brandi JamesAug 26, 2025
      If the system is in Safe Mode, a Self-Test is required before the system can be restored for normal operation. Click YES to perform Self-Test and resume work, or click NO to continue in Safe Mode.
  • K
    kevin41Aug 27, 2025
    What to do if Medoc PATHWAY Medical Equipment says water temperature is too high?
    • C
      Clifford GonzalezAug 27, 2025
      If the water temperature is too high: 1. Check that the PATHWAY is turned ON and communication is established. 2. Check that the Thermodes are properly connected. 3. Check the water level and fill the Cooling Unit reservoir if necessary. 4. Make sure the coolant temperature in the Cooling Unit is below 15°C. Press in the Emergency button and allow the temperature to cool if needed.
  • C
    Cody ShieldsSep 22, 2025
    What to do if Sensor Heater\TEC number of defined calibration table entries is less than the required 15 on Medoc PATHWAY Medical Equipment?
    • S
      Sydney AndersonSep 22, 2025
      If the calibration cannot be completed because there are not enough calibrated points, calibrate at least 15 points in order to complete the manual calibration.
  • V
    Victor ThomasSep 23, 2025
    What to do if Medoc PATHWAY shows Water overheat?
    • R
      Robert FordSep 23, 2025
      If the temperature of the coolant in the Thermode is too high: 1. Press the emergency button and wait 5-10 minutes to allow the coolant temperature to decrease. 2. Run the system Self-Test again. If the problem persists, there is no other solution.
  • F
    forbespatriciaSep 9, 2025
    What does the following safety limitation has been violated mean for Medoc PATHWAY Medical Equipment?
    • J
      jimenezrossSep 9, 2025
      If the temperature duration limit has been exceeded: 1. Check that the parameters of the program used are valid. 2. Check that the Thermodes are intact and undamaged. 3. Run the system Self-Test. If the problem persists, there is no other solution.
  • M
    morrisonnicoleSep 12, 2025
    What does [ATS / CHEPS] Thermode was not detected by Pathway mean for Medoc Medical Equipment?
    • T
      Todd WalkerSep 12, 2025
      If the system could not detect one of the enabled Thermodes, and therefore disabled it in order to prevent any damage: 1. Make sure that the Thermode is properly connected to the Pathway system. 2. Enable the Thermode through the Hardware Settings, General menu. 3. Click Yes to try again, or click No to continue working.
  • S
    Sara CummingsSep 3, 2025
    What to do if Medoc PATHWAY shows a dynamic sensor temperature mismatch?
    • M
      Melissa CareySep 3, 2025
      If the system has detected a mismatch between the Thermode sensors during a dynamic temperature change: 1. Check that the Thermodes are intact and undamaged. 2. Run the system Self-Test. 3. Re-calibrate the Thermode. If the problem persists, there is no other solution.
  • E
    emily92Sep 1, 2025
    What to do if Medoc PATHWAY Medical Equipment shows a static sensor mismatch?
    • M
      mjacobsSep 1, 2025
      If the system has detected a mismatch between the Thermode sensors: 1. Check that the Thermodes are intact and undamaged. 2. Run the system Self-Test. 3. Re-calibrate the Thermode. If the problem persists, there is no other solution.

Summary

Safety Guidelines and Regulations

Safety Requirements

Details general safety requirements for operating the PATHWAY system.

Warnings

Lists critical warnings to prevent serious injury, death, or equipment damage during operation.

Cautions

Provides precautions to prevent equipment damage or inaccurate system functioning.

Temperature Safety Mechanisms

Explains the safeguards implemented to protect against extreme temperatures.

Setup and Installation

Computer Requirements

Lists the minimum computer specifications required for Medoc Main Station software.

Software Installation

Details the procedure for installing the Medoc Main Station software.

Powering up the PATHWAY System

Provides step-by-step instructions for safely powering on the PATHWAY system.

Operation

Safety Procedures and System Self-Test

Details essential safety checks and the system self-test before starting operation.

Starting PATHWAY

Provides the sequence of steps to power on and initiate the PATHWAY system.

Test Procedure and Management

Running the Test

Detailed steps for initiating and conducting an actual test session.

Program Management

Creating a New Program

Step-by-step instructions for creating a new test program.

Results Management

Generate Test Reports

Steps for creating, previewing, and printing test reports.

System Configuration

Communication

Configuration settings for establishing communication between system and software.

Hardware Settings

Configuration options for hardware components like TTL, Thermodes, and CoVAS.

Appendix O - Troubleshooting

Frequent Warning Messages

Lists common warning messages and provides troubleshooting guidance.

Frequent Error Messages

Lists common error messages and provides troubleshooting guidance.

Faulty System Responses

Describes common system malfunctions and their corresponding solutions.

Medoc PATHWAY Specifications

General IconGeneral
CategoryMedical Equipment
Cooling RateUp to 8°C/sec
Heating RateUp to 8°C/sec
Power Supply100-240V AC, 50/60Hz
Stimulus TypesThermal
Temperature Range0°C to 55°C
InterfaceTouchscreen
ConnectivityUSB