Pathway Technical Reference Manual
Proprietary 52 of 190
8. When communication is established, the system will enter SAFE MODE and remain in this
state until completing the system Self-Test, see section 5.1 above. Make sure the
Thermode is not attached to the Subject and click on OK on the following message:
Figure 21: Self-Test Prompt
9. The system will perform a complete Self-Test of all enabled Thermodes according to your
system settings.
10. After the system Self-Test has completed successfully, the system state will change to
REST MODE.
Figure 22: PATHWAY in Rest Mode