EasyManua.ls Logo

Medoc PATHWAY - 22 Appendix J - Preventive and Periodical Maintenance; Summary Maintenance Table; Daily Preventative Maintenance Operations; Flow Rate Check

Default Icon
190 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Pathway Technical Reference Manual
Proprietary 172 of 190
22. Appendix J Preventive and Periodical
Maintenance
This chapter describes the preventative and periodical maintenance procedures required in
order to maintain good system performance and safety of use.
22.1 Summary Maintenance Table
The following table summarizes the required maintenance procedures.
Table 41: Summarizing Maintenance Table
Procedure
Frequency
Comments
Reference
Visual check
Every use
Depending on usage. Perform once a
month if system is not used often
22.2.1
Flow rate check
Every use
22.2.2
System run
1 month
Only if system is not being used
regularly.
22.3
Thermode Calibration
6 months
Requires a calibration kit. Some
study protocols may require
Thermode calibration more often
22.4
Cooling Unit maintenance
6 months
22.5
Ground conductivity Test
1 year
For MRI systems only
22.6
Thermode Maintenance
2 years
22.7
22.2 Daily Preventative Maintenance Operations
22.2.1 Visual Check
The visual check is important to ensure that all system components are intact and properly
connected before using the system.
The Visual check is required in order to make sure that there is no physical
damage to system components, particularly to the Thermodes.
1. Check all accessories and Thermodes visually.
1.1. Check that the accessory connectors are intact.
1.2. Check that the all the accessories’ bodies are intact.
1.3. Check that there is no visible damage to the Thermode heads. Check especially that
the Kapton® cover of the CHEP Thermode is intact and not peeled.
2. Make sure both Thermodes are properly connected the PATHWAY system. If your license
only includes one Thermode type (CHEPS or ATS), make sure to connect a bypass water
tube in place of the water connectors of the missing Thermode.
3. Connect all available Pathway accessories (Response Unit, Panic Button and etc.)
4. Make sure coolant level is between the minimum and maximum level (add coolant if
required, refer to section 19 for instructions).
22.2.2 Flow Rate Check
The flow rate is important to ensure the Thermodes get enough circulation of coolant for safety
and good performance. The required flow rate is approx. 37 LPH (indicator should be center

Table of Contents

Related product manuals