Pathway Technical Reference Manual
Proprietary 178 of 190
26. Appendix N – Environmental Conditions
The environment conditions for PATHWAY system during standard use are as follows:
An ambient temperature range of+18°C to+24 °C
A relative humidity range of 30% to 75%, including condensation
An atmospheric pressure range of 700 hPa to 1060 hPa