Pathway Technical Reference Manual
Proprietary 76 of 190
7.7.3 Search
In this method, the intensity of the stimulus changes according to the patient's response.
Using the response unit or CoVAS the patient can control whether the temperature will
increase or decrease and to what extent. For an example graph and an explanation of the
method, refer to section 2.8.3, page 35.
Figure 39: Program Menu – Search
The table below describes the program parameters available for Search method:
Table 16: Program Parameters
Specify the device which will be used to control the temperature:
CoVAS or Response Unit
Time before
sequence (sec)
The time interval before the first trial. Valid range: 0 to 600 (sec).
Temperature from which each new stimulation sequence begins.
ATS: 10°C – 45°C
CHEPS: 30°C – 45°C
Destination
temperature (°C)
Defines the destination temperature before the search starts. Valid
range is limited by safety limitations and depends on the duration
time.
Destination rate
(°C/sec)
Defines the temperature change rate from the baseline to destination
temperature.
CHEPS: 0.1 to 70°C/Sec
ATS: 0.1 to 8°C/Sec