13
Only use the recommended batteries.
Do not attempt to recharge non-rechargeable batteries.
Do not dispose of batteries in a fire.
Remove the batteries if the thermometer is not to be used for a long period of time.
5 – Cleaning, Care and Storage
It is important to protect the infrared lens from dirt and damage; it is very delicate.
Use a clean, soft cloth to clean the surface of the device and LCD. Do not use solvents or immerse
the device in water or other liquids.
Always keep the thermometer within the storage temperature range of -4.0 °F to 131.0 °F (-20.0 °
to 55.0 °C) and humidity range of ≤ 93% (non-condensing) as specified.
It is recommended to store the thermometer in a dry location free from dust.
Do not expose the thermometer to direct sunlight, high temperature or humidity or any other
extreme environmen tal conditions as it may aect the proper operation of the device.
If the ambient temperature changes too much, such as moving the thermometer from a cold room
to a warm room, allow the thermometer to adjust to the new temperature. This may take up to 30
minutes. Condensation may have formed on the infrared sensor.
6 – Disposal
Batteries should be disposed of in accordance with local regulations.
Do not dispose of the thermometer in the unsorted municipal waste stream. Enquire about options
Warning
Do not recharge, disassemble, or dispose of in a fire.
The typical service life for a set of new batteries is 2000 measurements with an operation time of 18
seconds per measurement.
If the device does not function after replacing the batteries:
Check for proper battery polarity.
Try a fresh set of batteries.
The unit may have locked up if the batteries were removed while the unit was powered. In
this case, remove the batteries, wait 30 seconds, and then reinsert the batteries.
for environmentally friendly and appropriate disposal. Take local regulations into account.
7 – Callibration
The thermometer is initially calibrated at the time of manufacture. If this thermometer is used according to
the instructions, periodic re-adjustment is not required. If at any time you question the accuracy of the
temperature measurements, please contact service personnel.