EasyManuals Logo

Megger BM11D User Manual

Megger BM11D
22 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #16 background imageLoading...
Page #16 background image
Figure 3. shows a problem which can occur. If one end of the
sample is grounded and this is required to be connected to the
positive terminal, then the negative (measurement) lead is
surrounded by a 5 kV field. This is likely to cause at least 1 nA of
unwanted leakage current, representing a 5 T resistance in
parallel with the sample under test.
Figure 3.
When taking measurements above 100 G therefore, the user
should where possible ground the Guard Lead as shown in
figure 2, otherwise parallel leakage paths may occur.
Stress Considerations
Measurement above 100 G
Figure 1. shows the stresses and subsequent leakages
which will occur between the test leads if neither is
connected to earth (ground). These leakages have
significant effect and occur through the air itself.
Figure 1.
Figure 2. shows the effect of connecting the guard lead to
the ground. This reduces the stray leakage into the negative
(measurement input) terminal considerably, but this
technique is only permissible if the item under test is
isolated from the ground. (Isolated means insulated by a
resistance of at least 5 M for the positive terminal or 10 k
for the negative terminal).
Figure 2.
16
BM11D/21 Application Notes 1/6/00 10:59 am Page 4

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Megger BM11D and is the answer not in the manual?

Megger BM11D Specifications

General IconGeneral
BrandMegger
ModelBM11D
CategoryTest Equipment
LanguageEnglish

Related product manuals