USER AND MAINTENANCE MANUAL
FUTURA PLUS
o1027a.eps

Table of Contents

Questions and Answers

  • C
    Courtney GlennAug 13, 2025
    What to do if the motor of Merivaara Medical Equipment does not work?
    • R
      Roger BlackAug 13, 2025
      If the motor isn't working, first check if the motor connection or the control unit connection has come loose and re-connect it to the control unit. Ensure the power lead is properly plugged into the wall socket. If these steps don't resolve the issue, it could be due to a blown fuse, a faulty limit switch, a fault in the motor, or the control unit current limit being exceeded due to overloading of motor.
  • C
    Catherine MaddenAug 17, 2025
    What to do if the hand-held control unit of Merivaara Medical Equipment does not work?
    • R
      raygordonAug 17, 2025
      If the hand-held control unit isn't working, first check if the hand-held control unit connection has come loose and re-connect it to the control unit. If this doesn't solve the problem, the wire or hand-held control unit might be damaged.
  • K
    Keith WilsonAug 20, 2025
    Why is Merivaara Medical Equipment mattress base not maintaining height?
    • R
      richardtaylorAug 20, 2025
      If the mattress base isn't maintaining its height, it could be due to a faulty valve, a faulty seal, or dirt in the hydraulic system. In such cases, pump replacement is necessary.
  • J
    Jay PaulAug 25, 2025
    How to fix Merivaara Medical Equipment mattress base that will not rise?
    • K
      Kristen FarmerAug 25, 2025
      If the mattress base won't rise, it could be due to a low oil level or air in the hydraulic system. Bleed pump to resolve this issue.
  • T
    Tracy CaseyAug 28, 2025
    How to fix Merivaara Medical Equipment mattress base that will not lower properly?
    • Q
      qmunozAug 28, 2025
      If the mattress base won't lower properly, it's likely due to air in the hydraulic system. Bleed pump to address this.
  • M
    mromeroSep 1, 2025
    What to do if the bed pulls to one side when pushing Merivaara Medical Equipment?
    • L
      Laura PowellSep 1, 2025
      If the bed pulls to one side when pushing, it's probably because a castor is sticking. Replace castor to fix this.
  • J
    Jeremy DaySep 5, 2025
    What causes Merivaara Medical Equipment mattress base angle adjusters to not remain in place?
    • N
      Nancy SandovalSep 5, 2025
      If the mattress base angle adjusters don't remain in place, it's likely because the gas spring is damaged or installed incorrectly. Replace gas spring to resolve this.
  • C
    Chase GarciaSep 9, 2025
    What to do if the function running does not correspond to the function button selected on Merivaara Medical Equipment?
    • C
      Cynthia JacksonSep 9, 2025
      If the function running doesn't match the selected function button, it means the motor leads are in the wrong order. Re-connect to control unit in numerical order to correct this.

Merivaara FUTURA PLUS Specifications

General IconGeneral
TypeOperating Table
ApplicationSurgical Procedures
Lateral Tilt20°
Height Adjustment650 - 950 mm
Table Width550 mm
Trendelenburg Tilt30°
Leg Section Angle90°
CertificationCE