2
Precautions
▪ Use only original components provided by the manufacturer.
▪ Leave a minimum space of 30 cm between the wall and the oven, to favor ventilation.
▪ To avoid the risk of electric shock, this equipment must only be connected to a suitable
power source (23 V, 50/60 Hz with earth connection).
▪ Use the oven between 5-40 ºC room temperature. Otherwise failures may occur.
▪ If there are shot residues or foreign substances in the chamber, remove them before use,
as they may cause furnace failure.
▪ Hairline cracks may occur in the insulating walls of the muffle and door. It is not a quality
problem nor does it affect the proper functioning of the oven. If in doubt, consult your
dealer.
▪ For safety, turn off and unplug the oven when not in use.
▪ If in doubt or if you need help, contact your dealer.
Warranty
All MESTRA® machines are guaranteed against any manufacturing defect for one year from the
day of purchase. To meet any guarantee, the purchase invoice will be required.
Failures caused by poor installation or misuse of the machine are excluded from the guarantees.
MESTRA® will not carry out any repair outside its workshop and is exempt from the transport
costs that this may cause. Repairs made at the customer's home, whether or not they are under
warranty, are entrusted to the technical service of the distributor that made the sale or to its
contracted technicians.
The warranty for the magnetron is 1 year or 300 cycles, although its useful life can reach 2
years or 400 cycles.