8
AC VOLTS (45 Hz - 500 Hz)
Ranges: 200m, 2 V, 20 V, 200 V, 600 V
Accuracy: All ranges, ± (1.5 % rdg + 4
dgts)
Resolution: 100 µV in 200m V range
Input impedance: 10 MΩ
Overload protection: 200mV range: 600
V dc or 600 V ac rms 15 seconds. Other
ranges: 600 V dc or 600 V ac rms
DC CURRENT
Ranges: 200µ A, 2m A, 20m A, 200m A,
10A
Accuracy:
200µ A to 200m A ranges: ± (1.5 % rdg
+ 1 dgt)
10A range: ± (2.0 % rdg + 3 dgts)
Resolution: 0.1 µA in 200µ A range
Burden voltage:
200 µA Range: 1 mV/ 1 µA
2 mA Range: 100 mV/ 1 mA
20 mA Range: 13 mV/ 1 mA
200 mA: 4.6 mV / 1 mA
10A: 40 mV / 1 A
Overload Protection:
µA / mA input: F 0.25A /600V, Min. I.R.
30 kA, (6.3x32 mm)
10A input: F 10A / 600V, Min. I.R.
100 kA, (10x38mm) (
10 A for 4 minutes
maximum followed by a 12 minute cooling
period
)
AC CURRENT (45 Hz - 500 Hz)
Ranges: 200µ A, 2m A, 20m A,
200m A,10A
Accuracy:
200µ A to 200m A ranges: ± (2.0 % rdg
+ 4 dgts)
10A range: ± (2.5 % rdg + 4 dgts)
Resolution: 0.1 µA in 200µ A range
Burden voltage: See DC Current
Overload Protection:
µA / mA input: F 0.25A /600V, Min. I.R.
30 kA, (6.3x32mm)
10A input: F 10A / 600V, Min. I.R.
100 kA, (10x38mm) (
10 A for 4 minutes
maximum followed by a 12 minute cooling
period
)
RESISTANCE
Ranges: 200 Ω, 2k Ω, 20k Ω, 200k Ω,
2M Ω, 20M Ω
Accuracy:
200 Ω to 200k Ω ranges: ± (1.0 % rdg
+ 4 dgts)
2M Ω ranges: ± (1.5 % rdg + 4 dgts)
20M Ω range: ± (2.0 % rdg + 5 dgts)
Resolution: 100 mΩ in 200 Ω range
Open circuit volts:
200 Ω range: 3.0 V dc
Other ranges: 0.3 V dc typical
Overload protection: 600 V dc or
600 V ac rms
CONTINUITY
Audible indication: 75 Ω ± 25Ω
Response time: 100 ms
Overload protection: 600 V dc or
600 V ac rms
DIODE TEST
Test current: 1.0 mA (approximate)
Accuracy: ± (1.5 % rdg + 3 dgts)
Resolution: 0.001 V
Open circuit volts: 3.0 V dc typical
Overload protection: 600 V dc or
600 V ac rms
BATTERY TEST
Ranges: 1.5V, 9V
Accuracy: ± (3.5 % rdg + 2 dgts)
Resolution: 1 mV, 10 mV
Load Test current:
1.5V range: 150 mA typical
9V range: 5 mA typical
Overload protection: 600 V dc or
600 V ac rms
NON-CONTACT VOLTAGE (NCV)
AC Volts: 70 V to 600 V ac
Red LED and Audible Indicator
REPLACEMENT PARTS
TL36 – Test Lead Set w/ Alligator clips
FP375 - Fuse Pack 250mA/600V
(4 each)
FP160 - Fuse Pack 10A/600V (2 each)