BCScale
SettinguptheScalesforEthernet:
1. SelectthecorrectSerialprotocol
setting.
Home
DefltDefFEnter
2. UpdateEthernetIPsetting
Home
SetupCOMMETH
PressEnter.
Note:AdefaultstaticIPsettinghas
beenprogrammedintothescale
PressUP/DOWNkeytoselect
IPaddressbyDHCP
(currentlynotsupported)
StaticIPaddress
OK
Cancel
3. UpdatestaticIPaddress:
HighlightthestaticIPaddressfield
PressEnter
key
PressRIGHT/LEFTkeytoscroll
throughtheoctetorENTERtoscroll
througheachIPaddress,subnet,
andgatewaysetting.
PressUP/DownkeytochangeIP
addressoctetvalues,example;
(192.168.000.002)
4. SaveupdatedstaticIPaddress:
Placecursortothefirstoctetofthe
IP
addressandpressLEFTkey.
PresstheDownkeytoOKthenpress
Enter



Other manuals for Mettler Toledo BC

Question and Answer IconNeed help?

Do you have a question about the Mettler Toledo BC and is the answer not in the manual?

Mettler Toledo BC Specifications

General IconGeneral
Capacity6 kg / 15 kg / 30 kg
InterfaceRS232
DisplayLCD
Units of Measurementg, kg, lb, oz
CalibrationExternal
MaterialStainless steel
Readability0.1 g / 0.5 g / 1 g
Protection ClassIP65 (dust and water resistant)

Summary

BC Scale Ethernet Setup

Select Serial Protocol Setting

Configure the scale's serial protocol setting via the menu.

Update Ethernet IP Settings

Modify the scale's IP address and subnet mask in the COMM menu.

Set Static IP Address

Manually enter a static IP address for the scale's network connection.

Save IP Address Configuration

Confirm and save the new IP address settings on the scale.

PC Network Connection Setup

Open Control Panel

Access the Windows Control Panel to manage network settings.

Access Network and Sharing Center

Navigate to the Network and Sharing Center for connection management.

Configure Local Area Connection

Select and open properties for the relevant Local Area Connection.

Set TCP/IPv4 Properties

Enter the scale's IP, Subnet Mask, and Gateway in TCP/IPv4 settings.

Close Network Windows

Close the connection status and properties windows after configuration.

Test Ethernet Connectivity

Perform Ping Test

Use Command Prompt to ping the scale's IP address to verify network reachability.

Use Serial Terminal Tool

Open a serial terminal, configure it for TCP/IP, and send commands.

Test Scale Serial Communication

Configure Scale Serial Protocol

Set the scale's serial protocol and COMM selection to ETH.

Configure Serial Tool Settings

Set the serial tool's Port Type, IP address, and Port to match the scale.