SA 35 and SA 36
Operation and Maintenance Manual
Manual 118 Issue 15
Page 23 of 28
Guidance and manufacturer’s declaration
electromagnetic emissions
It is important that the
SA
35/36 is configured and installed/put into service, in accordance with the
instructions/guidance provided herein and is used only in the configuration as supplied. For the purposes of
EN60601-1-2:2014, the
SA
35/36 has no essential performance.
It should be noted that cables / accessories provided with the
SA
35/36 should not be used on other equipment.
To do so may result in increased emissions or decreased immunity of the other equipment in relation to EMC
performance.
Emission test
Compliance
Electromagn
etic environment
-
guidance
RF emissions
CISPR 11
Group 1
The
SA
35/36 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
If the
SA
35/36 is used with cables / accessories other than those
supplied, this may result in increased emissions or decreased
immunity of the
SA
35/36 in relation to EMC performance.
The
SA
35/36 should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use with other equipment is
necessary, the
SA
35/36 and the other equipment should be
observed/monitored, to verify normal operation in the configuration
in which it will be used.
RF emissions
CISPR 11
Class B
The
SA
35/36 is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
If the
SA
35/36 is used with cables / accessories other than those
supplied, this may result in increased emissions or decreased
immunity of the
SA
35/36 in relation to EMC performance.
The
SA
35/36 should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use with other equipment is
necessary, the
SA
35/36 and the other equipment should be
observed/monitored, to verify normal operation in the configuration
in which it will be used.
Harmonic emissions
IEC61000-3-2
Class A
Voltage fluctuations /
flicker emissions
IEC61000-3-3
Complies
Guidance and manufacturer’s declaration
electromagnetic immunity
The purpose of this testing is to ensure the SAM 35/36 is not likely to adversely affect the normal operation of
other such equipment and that other such equipment is not likely to adversely affect the normal operation of the
SA
35/36.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environment - guidance
Electrostatic
discharge (ESD)
IEC60601-1-2:2014
IEC61000-4-2:2008
±2Kv, ±4kV,
±6kV and ±15kV
air discharge
±6kV contact
discharge
±2Kv, ±4kV,
±6kV and
±15kV air
discharge
±6kV contact
discharge
The
SA
35/36 should not be used adjacent to
or stacked with other equipment. If adjacent or
stacked use with other equipment is necessary,
the
SA
35/36 and the other equipment should
be observed/monitored, to verify normal
operation in the configuration in which it will be
used.
Electrical fast
transient / burst
IEC60601-1-2:2014
IEC61000-4-4:2012
± 2 kV for power
supply lines
± 1 kV for input /
output lines
± 2 kV for power
supply lines
Input / output
line tests not
applicable.
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC60601-1-2:2014
IEC61000-4-5:2005
Test level 1
±0.5kV, ±1.0kV
Test level 2
±2 kV
Test level 1
±0.5kV, ±1.0kV
Test level 2
±2 kV
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
Test level 1
Reduction of the
supply voltage of
100% for a half
and 1 periods
Test level 1
Reduction of the
supply voltage
of 100% for a
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the
SA
35/36 requires continued operation
during power mains interruptions, it is

Related product manuals