SA 35 and SA 36
Operation and Maintenance Manual
Manual 118 Issue 15
Page 24 of 28
IEC60601-1-2:2014
IEC61000-4-11:2004
Test level 2
Reduction of the
supply voltage of
70% for 25
periods
Test level 3
Reduction of the
supply voltage of
100% for 250
periods
half and 1
periods
Test level 2
Reduction of the
supply voltage
of 70% for 25
periods
Test level 3
Reduction of the
supply voltage
of 100% for 250
periods
recommended that the
SA
35/36 be powered
from an uninterruptible power supply or a
battery.
Power frequency
(50Hz)
Magnetic field
IEC60601-1-2:2014
IEC61000-4-8:2009
30 A/m 30 A/m
The
SA
35/36 should not be used adjacent to
or stacked with other equipment. If adjacent or
stacked use with other equipment is necessary,
the SAM 35/36 and the other equipment should
be observed/monitored, to verify normal
operation in the configuration in which it will be
used.
If incorrect operation occurs, it may be
necessary to position the
SA
35/36 further
from sources of power frequency magnetic
fields or to install magnetic shielding. The power
frequency magnetic field should be measured in
the intended installation location to assure that
it is sufficiently low.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration
electromagnetic immunity
The
SA
35/36 is intended for use in the electromagnetic environment specified below. The customer or the
user of the
SA
35/36 should assure that it is used in such an environment.
Immunity to RF test
Electromagnetic environment
-
guidance
Basic standard
IEC60601-1-2:2014 (table 9)
Test setup
IEC61000-4-3:2006;
Amd, 1:2007 and Amd, 2:2010
If the
SA
35/36 is used with cables / accessories other
than those supplied, this may result in increased
emissions or decreased immunity of the SAM 35/36 in
relation to EMC performance.
The
SA
35/36 should not be used adjacent to or
stacked with other equipment. If adjacent or stacked
use with other equipment is necessary, the
SA
35/36
and the other equipment should be
observed/monitored, to verify normal operation in the
configuration in which it will be used.
Despite the testing of the
SA
35/36 that has been
undertaken, normal operation of the
SA
35/36 can be
affected by other electrical/electronic equipment and
portable and mobile RF communications equipment.
Interference may occur in the vicinity of equipment
marked with the following symbol: