EasyManua.ls Logo

Micro Direct Spiro - 4 Zone Management Settings; Performing a Spirometry Test

Default Icon
10 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4
1. Turn the device on.
2. When the device is ready for a test , press the and
“Enter” buttons together for approximately 10 seconds.
3. The top (Green/Yellow) boundary can now be set. This is done
by pressing the or button and releasing when the value is
reached.
4. Press “Enter” to set the top (Green/Yellow) boundary value.
5. The bottom (Yellow/Red) boundary can now be set. This is
done by pressing the or button and releasing when the value
is reached. Press “Enter” to set the bottom (Yellow/Red) boundary
value.
6. Only 2 boundaries are required for the 3 zone system, so the
next value should be selected as 0% (default). Press “Enter”. The
device will return to the test screen.
Setting Management Zones (4 Zones)
1. First set the top and bottom boundaries see above procedure
(steps 1 - 6).
2. The middle (Yellow/Orange) boundary can now be set. This is
done by pressing the button and releasing when the middle
boundary value is reached. The values will increase/decrease in
values of 1% after an initial jump to the lower boundary value. If
the button is kept depressed, the values will scroll faster. This
boundary value cannot be set at a value that is greater than the
top boundary value or less than the bottom boundary value.
3. Press “Enter” to set the middle (Yellow/Orange) boundary
value. The device will return to the test screen.
Performing the Test