EasyManua.ls Logo

MICRO-EPSILON TIM 200 - Page 53

MICRO-EPSILON TIM 200
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 53
Appendix | Industrial Process Interface (Optional)
thermoIMAGER TIM
The industrial process interface offers the following inputs and outputs:
Designation Description max. range
1
/ status
A IN 1 / 2 Analog input 1 und 2 0 - 10 V
D IN 1 Digital input 24 V
AO1 / 2 / 3 Analog output 1, 2 and 3 0 - 10 V
Alarm output 1, 2 and 3 0 / 10 V
DO1 / 2 / 3 Relay output 1, 2 and 3
2
open/ closed (red LED on) / 0 ... 30 V, 400 mA
FS Fail-safe relay open/ closed (green LED on) / 0 ... 30 V, 400 mA
The process interface has an integrated fail-safe mode. This allows to control conditions like interruption of
cables, shut-down of the software etc. and to give out these conditions as an alarm.
Controlled conditions on camera
and software
Standard process interface
TM-PIF-TIM
Industrial process interface
TM-PIF500V2-TIM
Interruption USB cable to camera
Interruption data cable camera - PIF
Interruption power supply - PIF
Shut-down of TIM Connect software
Crash of TIM Connect software
1
-
Fail-safe output 0 V at analog output (AO) Open contact (Fail-safe relay)/
green LED off
1) Depending on supply voltage
2) Active if AO1,2 or 3 is/ are programmed as alarm output.