6 IM-5025 REV E
WARNING: The healthcare provider performing any procedure is responsible for determining whether this is the appropriate
equipment/instrument for the specic procedure that is being performed for each patient. As the manufacturer,
MicroAire does not recommend surgical procedures or oer any guidance in such procedures.
WARNING: Only trained and experienced healthcare providers/professionals should use this medical equipment. Before
using any system component or any component that is compatible with this system read and understand the
Instructions for Use. Pay special attention to the WARNING information and familiarize yourself with the system
components prior to use. Failure to comply may result in patient and/or healthcare professional injury.
WARNING: Consult and Follow the appropriate Instructions for Use for the instrument to be used. See the table under
Compatible Instruments.
WARNING: Do not use this equipment or components in the presence of ammable anesthetic mixture with air or with
oxygen or nitrous oxide.
WARNING: AC mains disconnect must be achieved by removing cord from wall outlet. Ensure that the Console is positioned
so that the power cord can be unplugged.
WARNING: Use personal protective equipment as required. Failure to comply may result in patient and/or healthcare
professional injury.
WARNING: Cleaning and Sterilization of all system components used inside the sterile eld are required before each use.
See MicroAires Cleaning and Sterilization Procedures in the appropriate instrument Instructions for Use.
WARNING: Use only MicroAire approved accessories. Use of unapproved accessories may result in increased
electromagnetic emissions or decreased immunity of the system and will void the warranty. Do not modify
any accessory. Failure to comply may result in patient and/or healthcare professional injury.
WARNING: Do not place instruments, components or handpieces on the patient. Failure to comply may result in
patient injury.
WARNING: Do not place handpieces near or on magnetic pads or trays. Magnetic elds can create throttle
output of a Foot Pedal or handpiece switch and may cause the instrument or component to operate
inadvertently. Failure to comply may result in patient and/or healthcare professional injury.
WARNING: Do not allow ngers to contact moving parts.
WARNING: This product does not comply with the ATEX directive and must
not be used with explosive atmospheres.
WARNING: Irrigation must be used when cutting bone in order to ensure that the temperature at the cutting
accessory does not exceed 41°C / 105.8°F.
WARNING: Use only with a medically approved type SJT or equivalent three wire (protective earth) grounded
power cord rated for 125V or 250V AC, 10A, constructed with 18/3 AWG wires and an IEC 320 appliance
connector.
WARNING: DO NOT modify the ground connection or any portion of the Console power cord. Failure to comply
may result in healthcare professional injury.
WARNING: Grounding reliability can only be achieved when the equipment is connected to an equipment
receptacle marked as “Hospital Grade”.
WARNING: Anyone who connects additional equipment to the contact closure connector congures a medical
system, and is therefore responsible that the system complies with the requirements of clause 16 of
ANSI/AAMI/IEC ES60601-1:2005 + Am1 (2012).
WARNING: The USB port is to be used only for diagnostics and/or software upgrade by MicroAire approved
technicians only. The USB port must not be connected to any cable or device while in operation.
WARNING: If the Console reboots due to instrument overload or loss of electrical power, the Irrigation Pump
will be disabled when the Console restarts. The Irrigation Pump must be re-enabled via the
touchscreen before resuming use of the instrument. Failure to comply may result in patient injury.
WARNING: Do not use this equipment in the presence of ammable anesthetics.
WARNING: Modication of this equipment is not allowed.
WARNING: Electric Shock. Do not remove cover. Refer servicing to authorized MicroAire personnel only.
WARNING: Take special precautions with electromagnetic compatibility (EMC) when using the MicroAire Electric Instrument
Console. Ensure that the Console is installed and placed into service according to the EMC information contained
in the Instructions for Use.