7IM-5025 REV E
WARNING: The 5020/5025 Electric Instrument Control Console and associated instruments have been tested for use in the
vicinity of High Frequency Electrosurgical/ Electrocoagulation equipment. Use of such equipment in close
proximity to the MicroAire instruments or instrument cables may cause unintended interruptions or unintended
motion of durations up to 1 second. Place MicroAire Instruments on a safe surface when not in use. Do not place
MicroAire Instruments on the patient when not in use. Avoid entwining HF surgical cables with MicroAire
Instrument Cables.
WARNING: Portable RF communications equipment should be used no closer than 30 cm (12 inches) away from any part of
the 5020/5025 Console, its associated cables or the handpieces it is controlling. Otherwise degradation of the
performance of the PAL-650 or series 5000 instruments could result.
WARNING: Use of the 5020/5025 Console adjacent to or stacked with other equipment should be avoided because it could
result in improper operation. If such use is necessary, the 5020/5025 and the adjacent equipment should be
observed to verify that they are operating normally.
WARNING: The following items should be periodically inspected for signs of damage and repaired or replaced as needed to
ensure continued safety with regard to electromagnetic disturbances over the life of the system:
• 5020/5025 Electric Instrument Control Console:
- Proper Fit of the top cover to the chassis.
- Damage to the LCD Display
- Test that the Protective Earth Connection meets ES60601-1 requirements.
• Inspect the 5006-PAL instrument cables for signs of physical damage to the cable.
WARNING: The 5020 & PAL-650 Liposuction handpiece and the 5025 & Series 5000 Instrument System are suitable for use in
hospitals and surgery centers. These systems should not be used near Magnetic Resonance Imaging equipment.
CAUTION: Federal Law (U.S.A.) restricts this device to sale by or on the order of a physician (or properly licensed
practitioner).
CAUTION: Do not thread or twist the push/pull connectors on the cable(s) during installation or removal.
CAUTION: When connecting or disconnecting a cable(s) to the front of the Console, always hold the cable by its
connector. Failure to comply may result in damage to the cable or Console.
CAUTION: Cables that are connected to the front of the Console have keyed, push/pull type connectors that lock
into place. Do not force a connector into a Console port. Each connector and port has an alignment
mark to indicate proper cable orientation.
CAUTION: DO NOT sterilize the 5020/5025 Console or its power cord, or the 5401 Foot Pedal
- They may be wiped down with a germicidal cleaner. Take precautions to avoid allowing any moisture
inside of any Console opening and power cord.
CAUTION: DO NOT use aerosol sprays of any kind directly on the screen display of the Console. Clean by wiping
with a soft, damp cloth and drying with a soft dry cloth.
NOTE: When the Instrument Slider Bar is being adjusted, the number shows the desired maximum speed.
When the instrument is in use (rotating), it will show the commanded speed of the instrument as a
percentage of its rated speed.
NOTE: If the Console screen is touched in more than one place at the same time, an unintended button
activation may occur. Please make sure the settings are adjusted as required before operating the
accessories/ instruments.
NOTE: The Instrument Direction icon only changes direction of the pencil grip instruments. For all other instruments
this icon serves as an indicator to the user of which direction the instrument is currently running. Pressing the
direction icon when these instruments are connected results in an error tone and will not change the direction
of that instrument.
NOTE: Connection of the Potential Equalization Connector is not required for compliance with IEC 60601-1
or IEC 60601-1-2.
NOTE: The emissions characteristics of this product make it suitable for use in industrial areas and hospitals (CISPR 11
Class A). If it is used in a residential environment (for which CISPR 11 Class B is normally required) this equipment
might not oer adequate protection to radio frequency communication services. The user might be required to
take mitigation measures, such as relocating or re-orienting the equipment.
NOTE: Repairs or alterations to MicroAire® products made by anyone other than MicroAire® or an Authorized
MicroAire® Repair Facility will void that products warranty, and the customer will be responsible for any costs
related to returning the product to working condition.
NOTE: The Irrigation Pump is not intended for use with the PAL-650.