Using rechargeable batteries
You can also operate this device using rechargeable batteries.
Only use «NiMH» type reusable batteries.
Batteriesmustberemovedandrechargedwhentheat
battery symbol appears. They should not remain inside the
device as they may become damaged (total discharge as a
result of low use of the device, even when switched off).
Always remove the rechargeable batteries if you do not
intend to use the device for a week or more.
Batteries cannot be charged in the blood pressure monitor.
Rechargebatteriesinanexternalchargerandobservethe
information regarding charging, care and durability.
7. Using a Mains Adapter
You can operate this device using the aponorm
®
by microlife
mainsadapter(DC6V,600mA).
Only use the aponorm
®
by microlife mains adapter
available as an original accessory appropriate for your
supply voltage.
Ensure that neither the mains adapter or the cable are
damaged.
1. Plug the adapter cable into the mains adapter socket
4
in
the blood pressure monitor.
2. Plug the adapter plug into the wall socket.
When the mains adapter is connected, no battery current is
consumed.
8. Error Messages
If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «ERR 3», is displayed.
* Please consult your doctor, if this or any other problem occurs
repeatedly.
If you think the results are unusual, please read through
the information in «Section 1.» carefully.
Error
«ERR 1»
«ERR 2»
«ERR 3»
«ERR 5»
«HI»
«LO»
Description
Signal too
weak
Error signal
No pressure
in the cuff
Abnormal
result
Pulse or cuff
pressure too
high
Pulse too
low
Potential cause and remedy
The pulse signals on the cuff are too
weak.Re-positionthecuffandrepeat
the measurement.*
During the measurement, error signals
were detected by the cuff, caused for
instance by movement or muscle
tension.Repeatthemeasurement,
keeping your arm still.
An adequate pressure cannot be gene
-
rated in the cuff. A leak may have
occurred.Checkthatthecuffiscor
-
rectly connected and is not too loose.
Replacethebatteriesifnecessary.
Repeatthemeasurement.
The measuring signals are inaccurate
and no result can therefore be dis
-
played.Readthroughthechecklistfor
performing reliable measurements and
then repeat the measurement.*
The pressure in the cuff is too high
(over300mmHg)ORthepulseistoo
high (over 200 beats per minute).
Relaxfor5minutesandrepeatthe
measurement.*
The pulse is too low (less than 40
beatsperminute).Repeatthemeasu
-
rement.*